PHOTOMETER 5010 V5+

Semi-automated clinical Biochemistry analyzer

Possible Applications:

Biochemistry tests on whole blood, serum, plasma, or urine samples.

 

 

 

Brochure / Website


 

TECHNICAL SPECIFICATIONS

Type Semi-automatic, single-beam filter photometer
Light Source LED with long lifetime
Wavelength 340 nm and 380 nm – 730 nm
Wavelength Selection Automatic via 9-position filter wheel :
6 standard interference filters: 340 nm, 405 nm, 492 nm, 546 nm, 578 nm and 623 nm;
3 positions for optional filter of choice
Photometric Range 0 – 2.5 A
Cuvette System Micro flow cell: 32 µl, 10 mm light path interchangeable with normal standard cuvettes (macro or semi-micro, disposable or special optical glass)
Temperature Control
  • Internal Peltier element, temperature variable, pre-adjusted to 25 °C, 30 °C and 37 °C
  • Equilibration time for aspirated reaction mixture to reach 37 °C from ambient temperature: 15 sec
Aspiration System Built-in peristaltic pump driven by stepper motor, programmable aspiration volume controlled by infrared light barrier
Sipping Volume
  • Minimum 250 µl, typically 500 µl up to 2000 µl
  • Separate setting of aspirate volume and wash volume
Operator Interface Touchscreen for direct functions and alphanumerical inputs
Data Presentation Graphic display: white characters or symbols, blue backlight, dimension 240 * 128 dots
Integrated Printer Thermal printer, 24 characters per line
Languages English and French / German / Indonesian / Russian / Spanish
Memory
  • General operating software can be updated by PC
  • Reagent open system with capacity for up to 231 programmable methods
  • Import of data by touchscreen or PC
  • Up to 50 non-linear calibration curves with max 20 sets of points can be stored
Data Logging Up to 1000 results can be saved in the memory
Signal Port
  • Serial port for connection to an external printer and/or PC
  • RS232 connector for barcode reader for an easy management of sample identities
Measurement Procedures
  • Absorbance
  • End Point with factor, standard or multiple standards, with or without reagent blank and/or sample blank
  • Bichromatic end point
  • Kinetics with standard, factor or multiple standards, with or without blank
  • Fixed time with standard, factor or multiple standards, with or without reagent blank
  • Turbidimetry
  • Single, double and triple determinations
  • Curve fitting for non-linear standard curves
  • Free hemoglobin in combination with optional interference filters
Quality control Up to 50 methods can be controlled with two control sera
Levey Jennings plot
Measuring Time
  • Kinetic: variable from 3 – 19 deltas, time per delta 4 – 255 sec in 1 sec steps
  • Fixed time: variable from 0 – 1800 sec in 1 sec steps
Delay Time Programmable from 0 – 1800 sec in 1 sec steps
Mains Supply Range: 100 VAC up to 240 VAC at 50/60 Hz
Dimensions Length 33 cm x width 34 cm x height 18 cm
Weight 5.0 kg

How to perform an optic adjustment with Photometer 5010 V5+